Navisphere Storage System Initialization Utility  

2008-11-19 17:08:25|  分类: 存储 |  标签: |举报 |字号 订阅

  下载LOFTER 我的照片书  |

Navisphere Storage System Initialization Utility

The Navisphere Storage System Initialization Utility is supported for both AX-Series and CX-Series storage systems. It lets you discover storage systems, set network parameters and, for AX-Series storage systems, create management user accounts for those storage systems. If the storage system is an AX150 ISCSI storage system,  use the utility to set network parameters for the storage system's iSCSI data ports.  You must initialize the storage system in order to register the server HBAs with the storage system and have access to the storage system's data.

Note  In this document, the term AX-Series storage systems refers to both AX100 and AX150-Series storage systems.

 

Before you start

Running the Storage System Initialization Utility

AX150-Series only

Setting iSCSI data port parameters (AX150-Series iSCSI storage systems only)

Uninitializing a storage system

Resetting initialization values

AX100-Series only

Uninitializing a storage system

Resetting initialization values

CX-Series only

Changing network parameters for CX-Series storage systems

 

What Next?

 

Before You Start

Note  For more details on the following storage system configurations, refer to one of the following:

  • Getting Started - AX100/AX100SC document that shipped with your AX100-Series storage systems

  • Getting Started - AX150/AX150SC document that shipped with your AX150-Series storage systems

  • Getting Started Guide that shipped with your CX-Series storage systems

 

Before you start the initialization operation, make sure that

  • You know the following:

    • Static IP addresses for the storage system SPs (example, http://123.45.6.7, and 123.45.6.8)

    • Subnet mask

    • Default gateway address.

    • CX-Series only - names you will assign to the storage system SPs.

  • All HBAs and drivers are installed in the servers connected to the storage system.

  • All cabling is correct.

  • The server on which you installed the initialization utility and the storage system are connected to the same network subnet.

Note  This is required for storage-system initialization only. Once you assign a IP address to the storage system, the server and storage system can be on different subnets.

  • The server and storage system are plugged into a standard ac power source and  turned on.

Back to Top

 

Running the Navisphere Storage System Initialization Utility

The Navisphere Storage System Initialization Utility lets you discover storage systems, set network parameters and create management user accounts (AX-Series only). You can run the utility from the server on which you install it or directly from the CD.

Note  We recommend that you install the initialization utility on at least one server with network connectivity to the storage system.

1

 

To start the Initialization wizard

From the server

Click Start > Programs > EMC > Navisphere > Navisphere Storage System Initialization Wizard

From the CD

From the main menu of the AX-Series Setup CD, click Run Utilities from the CD.

 

The wizard opens to the welcome screen. This screen lists the prerequisites for successfully using the wizard. Read the introductory text and click Next to begin the storage system discovery operation.

The discovery operation scans the subnet for storage systems. During the discovery process,  the discovery screen's graphic is animated. Once the discovery operation is complete, the wizard lists all uninitialized and initialized storage systems that it finds, using the storage system's serial number. You will find the serial number on the storage system in a location and format similar to the following (the letters preceding the serial number (ex. APM) may change).

AX-Series storage system

CX600/700 storage system

Navisphere Storage System Initialization Utility - lsxblog - lsxblog的博客

Navisphere Storage System Initialization Utility - lsxblog - lsxblog的博客

All other CX-Series storage systems

 

Navisphere Storage System Initialization Utility - lsxblog - lsxblog的博客

 

 

The discovery operation finds

AX-Series storage systems, running Navisphere Express, with connections to the server.

-or-

  • CX-Series storage systems running the FLARE Operating Environment software version 02.16.xxx.5.yyy or higher.

  • AX100-Series storage systems running Navisphere Manager and the FLARE Operating Environment version 02.19.xxx.5.yyy or higher, and/or AX150-Series storage systems running Navisphere Manager and the FLARE Operating Environment version 02.20.xxx.5.yyy or higher.

 Note  If no storage systems were discovered, verify the cable connections to these storage systems.

 

2

To initialize a storage system, select a storage system from the Uninitialized Systems list and click Next.

Note  You can  view the current settings of an initialized storage system by selecting it from the Initialized Systems list and then clicking Next.

3

In the Management Port Settings screen, set the following network parameters for the storage system 10/100 management ports:

 

Storage Processor A or Storage Processor B IP Address/ Name

Sets the IP address and name (CX-Series only) for SP A and SP B.

Name is user definable.

 

Subnet Mask

Sets the subnet mask associated with the LAN (local area network) for SP A and SP B.

 

Default Gateway

Sets the default gateway address used by the LAN for SP A and SP B.

4

For AX100-Series storage systems

For AX150-Series storage systems

For CX-Series storage systems - View the configuration summary

 

5

For AX150-Series iSCSI storage systems only, set the following iSCSI data port parameters, where X is the storage system storage processor - A or B, and Y is the SP's iSCSI data port number - 0 or 1.

Note  For AX100-Series iSCSI storage systems, you must use Navisphere Express to set the iSCSI data port parameters.

 

SP X Port Y  IP Address

Sets the IP address for  the iSCSI data port.

 

SP X Port Y Subnet Mask

Sets the subnet mask for the iSCSI data port.

 

SP X Port Y Direct Attach

Select Direct Attach if this port connects directly to the server and  not through a LAN (local area network).

This sets the port's gateway address to 0.0.0.0 and makes the gateway parameter unavailable for editing.

 

SP X Port Y Gateway

Sets the gateway address for the  iSCSI data port.

 

Click Next to change the storage system's name and set the storage system's login credentials.

6

For AX-Series (AX100 and AX150-Series) storage systems, in Storage System Name, enter a new name for the storage system.

7

For AX-Series storage systems, enter the following login credentials for the storage system.

Note  Security Settings entries are available only for AX-Series storage systems. For CX-Series storage systems,  use Navisphere Manager to create a domain and set domain security. CX-Series storage systems still display as uninitialized until you set domain security.

 

User name

Sets a user name. A valid name must start with a letter, can contain only alphanumeric characters, cannot exceed 32 characters, and is case sensitive.
For example, blindmice3 is a valid user name. Do not enter BLINDMICE3 by mistake.

 

Password

Sets the user password. A valid password can contain only alphanumeric characters, cannot exceed 32 characters, and is case sensitive. For example, mousetrap2 is a valid password. Do not enter MOUSETRAP2 by mistake.

 

Confirm password

Re-enter your password for verification.

8

Click Next to view the Configuration Confirmation screen. This screen summarizes the network settings selected for this storage system, and, for CX-Series storage systems, lets you change the name of an uninitialized storage system.

9

If the settings are correct, click Finish. If the settings are not correct, click Back until you reach the screen where you can modify the settings, and then proceed through the wizard screens again.

Note Once you click Finish, the storage system reboots and the SP Boot/Fault light on the back of each SP starts blinking. The reboot operation could take several minutes to complete. When it is completes, the SP Boot/Fault light on each SP stops blinking and remains off.

10

Once the initialization operation is complete, the wizard displays a results screen. This screen lets you

  • Create a desktop shortcut that will open the storage management tool (Navisphere Manager or Navisphere Express)

  • Initialize another storage system (selecting this check box brings you back to the screen that lists all uninitialized storage systems).

Back to Top

 

(AX150-Series Only) Resetting Initialization Values and Uninitializing the Storage System

Note Applies to AX150-Series storage systems only.

In order to reset any of the storage system's initialization values - username, user password, IP address, subnet mask or gateway address, you must uninitialize the storage system before you reset

Important  Each time you uninitialize the storage system, you must reset all security settings (username and password). The storage system retains its IP addresses, the subnet mask and the default gateway address.

To uninitialize the storage system and reset the values, do the following:

1

Power down the storage system by pressing the power button located on the back panel of the storage system.

Wait one minute before powering up again.

2

To restart the storage system, press and hold the power button (located on the back panel of the storage system) for 4 to 8 seconds, and then release it.

3

The storage system will proceed through three levels of start-up. You can differentiate between the levels by the flash rate of the Boot/Fault light located next to the reset button.

  • Flashes once every four seconds - BIOS activity

  • Flashes once per second - POST activity

  • Flashes flour times per second - booting

Once the start-up is complete, the storage system is uninitialized. Rerun the initialization utility to reset the initialization values.

Back to Top

 

(AX100-Series Only) Resetting Initialization Values and Uninitializing the Storage System

Note Applies to AX-Series storage systems only.

In order to reset any of the storage system's initialization values - username, user password, IP address, subnet mask or gateway address, you must uninitialize the storage system before you reset

Important  Each time you uninitialize the storage system, you must reset all security settings (username and password). The storage system retains its IP addresses, the subnet mask and the default gateway address.

To uninitialize the storage system and reset the values, do the following:

1

Power down the storage system by pressing the power button located on the back panel of the storage system.

Wait one minute before powering up again.

2

Restart the storage system by pressing the same power button and holding it for one second after the green power light comes on.

The storage system will proceed through three levels of start-up. You can differentiate between the levels by the flash rate of the Boot/Fault light located next to the reset button.

  • Flashes once every four seconds - BIOS activity

  • Flashes once per second - POST activity

  • Flashes flour times per second - booting

Once the start-up is complete, the storage system is uninitialized. Rerun the initialization utility to reset the initialization values.

Back to Top

 

Changing Network Parameters for CX-Series Storage Systems

Note This does not apply to AX-Series storage systems.

You can change the network parameters for a CX-Series storage system as many times as you want prior to adding the storage system to a domain and configuring domain security.  

Important  If you want to change the network parameters after configuring domain security, you must remove (destroy) the domain and the domain security (using Navisphere Manager), and then use the setup page to restart the management server software for both SPs. Open the setup page by entering the following URL in your web browser - SP IP Address/setup.
For example, 111.222.33.444/setup.

Back to Top

 

What Next?

In order to access the data stored on the storage system, you must register the HBAs of any attached servers with the storage systems.

For AX-Series storage systems - Use the Navisphere Server Utility that ships on the same CD as the Navisphere Storage System Initialization Utility.

For CX-Series storage systems - Reboot the server or use the Navisphere Server Utility.

Note Before running the Navisphere Server Utility, wait for the storage system to finish rebooting.

Back to Top